ICU Ventilator S1200 for Adult and Child

  • 15” TFT LCD touch screen displays the ventilation parameters, alarm information and waveforms.
  • Air compressor runs stably with low noise, creates quiet environment for doctors and patients.
  • Integration breathing circuit design.
  • Multiple working modes such as volume control and pressure limit.
  • Multiple parameters monitoring interface to show clear parameter to the operator, allow operator know the patient conditions in all aspects.
  • Real time pressure-time, flow-time loop graphics, pressure volume loop graphics and high precision ETCO2, O2 concentration detection function included.
  • Electrical PEEP control
  • Breathing valve is disassembly and can be disinfected.
  • Application

    Ybo ventilator is suitable for various kinds of medical institutions for cardiopulmonary resuscitation respiratory support. Acute respiratory failure caused by various reasons or incomplete oxygenation dysfunction. Intra-operation, post-operation respiratory support, other respiratory treatment.

    Safety
  • Three level alarming system, visual and sound alarm information.
  • With multiple type of alarms, reminding and protection functions.
  • Self-check before running, eliminate system mulfuction.
  • Advanced power management control technology.
  • With built-in backup power supply, when normal power supplyshut down, back-up power supplywill start work automatically.
  • Separate design of electric and gas, keep safety running of ventilator.
  • Specification
    Ventilation modes: IPPV, A/C, PCV, P-SIMV-PS, V-SIMV-PS,APRV, PRVC, SPONT/CPAP, PSV, Duolevel
    Ventilator parameter ranges  
    Tidal volume(Vt) ~ 2000mL
    Frequency (Freq) Oxygen concentration 1bpm ~ 100 bpm 21%-100%
    I/E 4:1~1:8
    PEEP 0cmH2O ~ 40 cmH2O
    Pressure Limit 20 cmH2O ~ 100 cmH2O
    Monitored Parameters  
    Frequency (Freq) 0 /min ~ 100 /min
    Tidal volume(Vt) 0 mL ~ 2500 mL
    MV Airway pressure Dynamic lung compliance testing ETCO2 Concentration 0 L/min ~ 99 L/min 0 cmH2O ~ 100 cmH2O 1mL/cmH2O ~ 1000mL/cmH2O 0mmHg ~ 152mmHg (0%-20%)
    Oxygen concentration 15 % ~ 100 % 
    Oscillographs display:  
    P-T(pressure – time)  
    F-T(flow – time)  
    P-V loop (pressure – volume loop)  
    V-T(volume – time )  
    ETCO2-T (End-tidal CO2 - time)  
       
    Alarms and protection  
    The AC power failure alarm Power failure or no connection
    Internal battery backup low voltage alarm 11.3±0.3V
    No tidal volume No tidal volume within 6s
    High Minute Volume alarm Low Minute Volume alarm 5L/min-99L/min 1L/min-30L/min
    High Airway pressure alarm Low Airway pressure alarm High oxygen concentration alarm Low oxygen concentration alarm 20cmH2O-100cmH2O 0cmH2O-20cmH2O 19%-100% 18%-99%
    Continuous pressure alarm (PEEP+1.5cmH2O) over 16s
    Suffocation warning Fan error Oxygen deficit                 5-60s Show on screen Show on screen
    The maximum limited pressure <12.5 kPa
       
    Working conditions  
    Gas source O2, AIR
    Pressure 280kPa-600kPa
    Voltage ~220V±22V
    Power frequency 50Hz ± 1Hz
    Input power 900VA (with air compressor) 250VA (without air compressor)